Write short notes on/answer the following in about 150 words each :
Use of electron microscopy in the study of tissue structure and cytology
10 marks · answer in 150 words · directive: Write short notes
The core of a good answer
Electron microscopes use a beam of electrons instead of light and resolve about 0.1 nm against about 200 nm for light microscopes; transmission EM of ultrathin, heavy metal stained sections shows the internal structure of cells and tissues, scanning EM of dried, gold coated specimens shows surface topography, and cryo methods examine vitrified specimens, but samples must withstand a high vacuum and can be altered by preparation and by the beam.
An outline written by upsc.help, not an official UPSC answer; every fact checked by a separate review. How we work
What the question asks
Explain the principle and types of electron microscopy, how tissues are prepared for it, what each type reveals about cell and tissue structure, and its limitations.
Answer outline
Introduction
In 1931 Max Knoll and Ernst Ruska produced magnified images with an electron beam, and in 1933 they built the first electron microscope that exceeded the resolution of a light microscope. Electron microscopes resolve about 0.1 nm, against about 200 nm for light microscopes.
Types
- Transmission electron microscope (TEM): electrons, generally of 80 to 300 keV, pass through a thin specimen and form an image of its internal structure; high resolution TEM resolves below 50 picometres.
- Scanning electron microscope (SEM): a beam generally below 20 keV scans the surface; the secondary electron signal is highly localised, and the large depth of field gives micrographs a characteristic three dimensional appearance.
- Cryo electron microscopy: since the 1980s, cryofixed, vitrified specimens held at liquid nitrogen temperature have been used increasingly.
Preparation of tissue for TEM
- Biological specimens are embedded in resin so that they withstand the high vacuum and can be cut thin.
- An ultramicrotome cuts sections of less than 100 nm as the resin block passes over a glass or diamond knife.
- Compounds of heavy metals such as osmium, lead and uranium give contrast by scattering electrons; gold is used in immunogold labelling to localise specific molecules.
- Negative staining with uranyl acetate outlines bacteria and viruses.
Preparation of tissue for SEM
- Fixation in a buffered chemical fixative such as glutaraldehyde.
- Water is replaced by organic solvents and then a transitional fluid and removed by critical point drying, because air drying causes collapse and shrinkage.
- The specimen is mounted on a stub and sputter coated with gold or gold palladium alloy to make it conductive.
Use in histology and cytology
- TEM of thin sections reveals the internal ultrastructure of cells and tissues far below the limit of the light microscope.
- SEM shows the surface topography of cells and tissues in three dimensional view.
- Immunogold labelling localises particular molecules within cells; negative staining identifies bacteria and virus particles.
Limitations
- The specimen chamber is under high vacuum, so specimens must be completely dry, embedded or cryogenically cooled.
- The electron beam can damage biological material, and preparation can change the structure of the sample (artefacts).
Conclusion
By resolving structures about two thousand times finer than the light microscope, electron microscopy turned the cell into a detailed map of its internal and surface structure, provided preparation artefacts and beam damage are kept in mind.
Examples and data to cite
- Ernst Ruska received the 1986 Nobel Prize for the invention of the electron microscope; Max Knoll, who died in 1969, did not share it.
- Negative staining with uranyl acetate to visualise bacteria and viruses.
Facts to remember
- Resolution of the light microscope is about 200 nm; of the electron microscope about 0.1 nm.
- TEM generally uses 80 to 300 keV electrons; SEM usually below 20 keV.
- Knoll and Ruska produced electron images in 1931 and built the first electron microscope to exceed light microscope resolution in 1933.
- Thin sections for TEM are under 100 nm thick.